Monopolar electrosurgery relies on extreme current density differences. At the surgical active pencil electrode, the contact area is tiny (<1 mm²), generating intense local heat that vaporizes intracellular fluid (Cut mode) or dehydrates tissue (Coag mode).
To complete the circuit safely, this exact same total current must exit the body through the patient return electrode. Because the return plate has a huge surface area (>100 cm²), the current density is diluted to harmless levels without heating.
However, if the return pad peels off by 60%, the remaining contact area suffers severe current concentration, causing 3rd-degree burns.