Because electrosurgical frequencies (300 kHz – 3.3 MHz) readily pass through stray capacitances in cables, the generator must have internal RF isolation. Under IEC 60601-2-2, high-frequency leakage to earth from both active and return leads must be measured into a 200Ω non-inductive load and must not exceed 150 mA.