Traditional ventilators utilized passive spring-loaded peep valves. Modern life-support ventilators utilize dynamic voice-coil or proportional solenoid-driven silicone diaphragm valves.
The microcontroller samples airway pressure at 1,000 Hz (every 1 millisecond). When the patient exhales, the coil current is modulated via Pulse-Width Modulation (PWM) to provide exact counter-force against the diaphragm. This maintains steady Positive End-Expiratory Pressure (PEEP) without pressure spikes, even when the patient coughs or moves.