1. Type B (Body): Applied parts intended for external contact or non-conductive internal contact, where the patient does not have direct cardiac access. Generally bonded to protective earth. Examples: Hospital motorized beds, phototherapy units, examination lamps.
2. Type BF (Body Floating): Applied parts with conductive patient contact that are electrically floating and separated from all other parts of the equipment. If external mains voltage accidentally touches the patient from another source, the floating circuit prevents dangerous currents from returning through the equipment to earth. Examples: Diagnostic ECG electrodes, ultrasound transducers, NIBP cuffs, pulse oximeter probes.
3. Type CF (Cardiac Floating): Applied parts designed for DIRECT CARDIAC CONTACT. They must provide the highest degree of galvanic isolation. Leakage currents are restricted to microscopic levels (10 µA under Normal Condition) to prevent inducing ventricular fibrillation. Examples: Intracardiac ECG leads, catheter ablation electrodes, cardiac pacing catheters, invasive blood pressure (IBP) transducer domes.
IEC 60601-1 Allowable Leakage Current Limits (RMS AC)| Leakage Current Type | Type B Applied Part | Type BF Applied Part | Type CF (Cardiac Floating) |
|---|
| Earth Leakage (Normal Condition) | 500 µA (0.5 mA) | 500 µA (0.5 mA) | 500 µA (0.5 mA) |
| Earth Leakage (Single Fault Condition) | 1000 µA (1.0 mA) | 1000 µA (1.0 mA) | 1000 µA (1.0 mA) |
| Touch / Chassis Leakage (NC) | 100 µA (0.1 mA) | 100 µA (0.1 mA) | 100 µA (0.1 mA) |
| Touch / Chassis Leakage (SFC) | 500 µA (0.5 mA) | 500 µA (0.5 mA) | 500 µA (0.5 mA) |
| Patient Leakage Current (NC) | 100 µA | 100 µA | 10 µA (0.01 mA) |
| Patient Leakage Current (SFC) | 500 µA | 500 µA | 50 µA (0.05 mA) |